Table 2:

Practical characteristics and performance of the 3 screening tests

Screening testTime taken for test, s, mean ± SDScreening test result no. (%)True positive (atrial fibrillation confirmed)*False positive (atrial fibrillation excluded)*Positive predictive value % (95% CI)Relative false positive rate compared with pulse-check
Pos.Neg.
Pulse-check
n = 2054
42.1 ± 22.5118 (5.7)1936 (94.3)456540.9 (31.7–50.1)
Single-lead
ECG
n = 2054
128.2 ± 82.078 (3.8)1976 (96.2)511873.9 (63.6–84.3)0.28
Blood pressure
n = 2054
194.8 ± 71.579 (3.8)1975 (96.2)393453.4 (42.0–64.9)0.52
  • Note: CI = confidence interval, ECG = electrocardiogram, Neg. = negative, Pos. = positive, SD = standard deviation.

  • * Numbers refer to all patients who underwent each test who attended confirmatory testing (12-lead ECG ± Holter monitor), so are less than the number screening positive in each case.