Table 1: Water and electricity consumption and cost per treatment for various combinations of treatment duration and dialysate flows (not including cost of intermittent chemical disinfection)
DurationWaterElectricityTotal cost per treatment,  $
Per treatment dialysate desired, LPer treatment water consumption, mean ± SD, m3 at 20°C
n = 5
% Dialysate desired to water consumedPer treatment water cost*,  $Per treatment electricity consumption, mean ± SD, kWh at 20°C 
n = 5
Per treatment electricity cost†, $
6 h × 300 mL/min1080.8052 ± 0.0390132.614.25 ± 0.460.382.99
8 h × 300 mL/min1441.0128 ± 0.0478143.295.16 ± 0.480.463.74
4 h × 500 mL/min1200.6675 ± 0.0328182.173.68 ± 0.220.332.49
6 h × 500 mL/min1800.8987 ± 0.0216202.924.70 ± 0.280.423.33
2 h × 800 mL/min960.4829 ± 0.0168201.572.64 ± 0.230.231.80
3 h × 800 mL/min1440.6047 ± 0.0288241.963.40 ± 0.260.302.26
4 h × 800 mL/min1920.7344 ± 0.0244262.384.04 ± 0.380.362.74

Note: SD = standard deviation.

*Assumes a water cost of $3.2438/m3 and includes a 60-min pretreatment machine warm-up and operations check in addition to a 45-min posttreatment heat disinfection cycle.

†Assumes an electricity cost of $0.089/kWh and includes a 60-min pretreatment machine warm-up and operations check in addition to a 45-min posttreatment heat disinfection cycle.